Assessment of left ventricular function by breath-hold cine MR imaging: Comparison of different steady-state free precession sequences
Purpose To compare steady-state free precession (SSFP) sequence protocols with different acquisition times (TA) and temporal resolutions (tRes) due to the implementation of a view sharing technique called shared phases for the assessment of left ventricular (LV) function by breath-hold cine magnetic resonance (MR) imaging. Materials and Methods End-diastolic and end-systolic volumes (EDV, ESV) were measured in contiguous short-axis slices with a thickness of 8 mm acquired in 10 healthy male volunteers. The following true fast imaging with steady-state precession (TrueFISP) sequence protocols were compared: protocol A) internal standard of reference, segmented: tRes 34.5 msec, TA 18 beats pe…