Optical Studies of the Capillary Refill Kinetics in Fingertips
A new methodology for temporal analysis of the capillary refill processes by means of photoplethysmography (PPG) contact probe operating in the blue region of spectrum has been developed. A corresponding prototype device for finger measurements has been created and clinically tested. Results demonstrated that both AC and DC components of the blue PPG bio-signals are sensitive to the capillary occlusion and refill. Real-time measurements of the skin capillary refill kinetics by this technology can be used for noninvasive diagnostics of the peripheral perfusion disorders.
Real-time analysis of skin capillary-refill processes using blue LED
A method for analysis of skin capillary-refill processes in real time by means of reflection photoplethysmography (PPG) contact probe operating in the blue (438nm ± 30 nm) and infrared (938 nm ± 20 nm) regions of spectrum is proposed. The corresponding prototype hardware and software for measurements have been developed and tested in laboratory. Realtime measurements of finger capillary refill kinetics by this technology have been taken and analyzed. Results demonstrated that both AC and DC components of the blue PPG biosignal are sensitive to capillary occlusion and refill.