Fast Response Heating Module for Temperature Programmed GC Analysis in Microreaction Systems
For monitoring the reactant and product concentrations of a gas phase microreactor for ethylene oxide synthesis the development of a miniaturised device for fast GC analysis suitable for operation in conjunction with the microreactor is advantageous. Therefore, a method employing short (0.2 – 0.4 m) packed capillary columns was developed. Highly crosslinked polystyrene microspheres proved to be the most suitable adsorbent enabling a high separation efficiency. For fast temperature programming design concepts for an efficient heating module capable to be used with short capillary columns were developed. In principle, planar as well as cylindrical designs are suitable for integration of a res…