ADDISC lumbar disc prosthesis: Analytical and FEA testing of novel implants

The intact intervertebral disc is a six-freedom degree elastic deformation structure with shock absorption. 'Ball-and-socket' TDR do not reproduce these properties inducing zygapophyseal joint overload. Elastomeric TDRs reproduce better normal disc kinematics, but repeated core deformation causes its degeneration. We aimed to create a new TDR (ADDISC) reproducing healthy disc features. We designed TDR, analyzed (Finite Element Analysis), and measured every 500,000 cycles for 10 million cycles of the flexion-extension, lateral bending, and axial rotation cyclic compression bench-testing. In the inlay case, we weighted it and measured its deformation. ADDISC has two semi-spherical articular s…

research product

Selective T3–T4 sympathicotomy versus gray ramicotomy on outcome and quality of life in hyperhidrosis patients: a randomized clinical trial

Abstract Background: compensatory hyperhidrosis is the leading cause of patients' dissatisfaction after thoracic sympathicotomy.Objective: to reduce compensatory hyperhidrosis to increase patients' satisfaction. Patients and methods: a prospective randomized study on palmar hyperhidrosis, May 2016-September, 2019. Twenty-one patients T3-T4 sympathicotomy and 21 T3-T4 gray ramicotomy. Data prospectively collected. Analysis at study's end. Focus on the sweating, temperature, quality of life baseline and postoperatively, compensatory hyperhidrosis, hand dryness, patients' satisfaction, and if they would undergo the procedure again and would recommend it.Results: No baseline differences between…

research product

Finite element analysis of a Bionate ring-shaped customized lumbar disc nucleus prosthesis

[EN] Study design: Biomechanical study of a nucleus replacement with a finite element model. Objective: To validate a Bionate 80A ring-shaped nucleus replacement. Methods: The ANSYS lumbar spine model made from lumbar spine X-rays and magnetic resonance images obtained from cadaveric spine specimens were used. All materials were assumed homogeneous, isotropic, and linearly elastic. We studied three options: intact spine, nucleotomy, and nucleus implant. Two loading conditions were evaluated at L-3-L-4, L-4-L-5, and L-5-S-1 discs: a 1000 N axial compression load and this load after the addition of 8 Nm flexion moment in the sagittal plane plus 8 Nm axial rotation torque. Results: Maximum nuc…

research product

Bionate biocompatibility: in vivo study in rabbits

Response to foreign materials includes local tissue reaction, osteolysis, implant loosening, and migration to lymph nodes and organs. Bionate 80A human explants show minor wear and slight local tissue reaction, but we do not know the response at the spinal cord, nerve roots, lymph nodes, or distant organs. This study aims to figure out reactions against Bionate 80A when implanted at the spinal epidural space of 24 20-week-old New Zealand white rabbits. In one group of 12 rabbits, we implanted Bionate 80A on the spinal epidural space, and another group of 12 rabbits was used as the control group. We studied tissues, organs, and tissue damage markers on blood biochemistry, urine tests, and ne…

research product

Bionate Lumbar Disc Nucleus Prosthesis: Biomechanical Studies in Cadaveric Human Spines.

Design: cadaveric spine nucleus replacement study. Objective: determining Bionate 80A nucleus replacement biomechanics in cadaveric spines. Methods: in cold preserved spines, with ligaments and discs intact, and no muscles, L3-L4, L4-L5, and L5-S1 nucleus implantation was done. Differences between customized and overdimensioned implants were compared. Flexion, extension, lateral bending, and torsion were measured in the intact spine, nucleotomy, and nucleus implantation specimens. Increasing load or bending moment was applied four times at 2, 4, 6, and 8 Nm, twice in increasing mode and twice in decreasing mode. Spine motion was recorded using stereophotogrammetry. Expulsion tests: cyclic c…

research product

Barbed Dental Ti6Al4V Alloy Screw : Design and Bench Testing

Background context. Dental implants are designed to replace a missing tooth. Implant stability is vital to achieving osseointegration and successful implantation. Although there are many implants available on the market, there is room for improvement. Purpose. We describe a new dental implant with improved primary stability features. Study design. Lab bench test studies. Methods. We evaluated the new implant using static and flexion–compression fatigue tests with compression loads, 35 Ncm tightening torque, displacement control, 0.01 mm/s actuator movement speed, and 9–10 Hz load application frequency, obtaining a cyclic load diagram. We applied variable cyclic loadings of prede…

research product

ICR in human cadaveric specimens: An essential parameter to consider in a new lumbar disc prosthesis design.

[EN] Study design Biomechanical study in cadaveric specimens. Background The commercially available lumbar disc prostheses do not reproduce the intact disc's Instantaneous centre of Rotation (ICR), thus inducing an overload on adjacent anatomical structures, promoting secondary degeneration. Aim To examine biomechanical testing of cadaveric lumbar spine specimens in order to evaluate and define the ICR of intact lumbar discs. Material and Methods Twelve cold preserved fresh human cadaveric lumbosacral spine specimens were subjected to computerized tomography (CT), magnetic resonance imaging (MRI) and biomechanical testing. Kinematic studies were performed to analyse range of movements in or…

research product

Nucleus disc replacement : designs and material selection FEA analysis

Study Design. We selected the materials and implant design and performed Finite Element Analysis (FEA) studies. Background. Nucleus disc replacements, implanted since 1960, have undergone continuous evolution in materials and designs, but subsidence, extrusion, and in vivo degradation limit widespread use. Aim. The aim of this study is to create a new nucleus disc replacement that avoids the abovementioned drawbacks. Material and Methods. We created eighteen designs with varied materials and analyzed them with FEA in compression and shear tests in a lumbar spine model programmed in Ansys Parametric Design Language. Results. Bionate® 80A had the closest mechanical characteristics to the inta…

research product