Dosimetric characteristics of backscattered electrons in lead.
In electron beam therapy, tissue overdose due to electrons backscattered from lead has been profusely studied. To quantify this dose enhancement effect, an electron backscatter factor (EBF) was defined as the ratio of dose at the tissue-inhomogeneity interface with and without the scatterer present. The dependence of the EBF on energy at the scatterer surface is not well known for energies lower than 3 MeV which is the most frequent clinical situation. In this work, we have done Monte Carlo calculations with the GEANT code to study EBF in lead at this energy range. The applicability of this code and the developed procedure for dose estimation has been experimentally verified. The dependence…