0000000000277511

AUTHOR

F. Celada

Novel simple templates for reproducible positioning of skin applicators in brachytherapy.

Purpose : Esteya and Valencia surface applicators are designed to treat skin tumors using brachytherapy. In clinical practice, in order to avoid errors that may affect the treatment outcome, there are two issues that need to be carefully addressed. First, the selected applicator for the treatment should provide adequate margin for the target, and second, the applicator has to be precisely positioned before each treatment fraction. In this work, we describe the development and use of a new acrylic templates named Template La Fe-ITIC. They have been designed specifically to help the clinical user in the selection of the correct applicator, and to assist the medical staff in reproducing the po…

research product

EP-1613: Depth determination of skin cancers treated with superficial barchytherapy: ultrasound vs. histopathology

research product

EP-1590: Shielding design and fetal dose evaluation of a breast cancer pregnant patient undergoing HDR Ir-192 brachytherapy

research product

Assessing the Relationship Between Dose-Volume Histogram Parameters and Late Rectal Toxicity in HDR Brachytherapy for Prostate Cancer

research product

Failure mode and effects analysis of skin electronic brachytherapy using Esteya® unit

Purpose: Esteya® (Nucletron, an Elekta company, Elekta AB, Stockholm, Sweden) is an electronic brachytherapy device used for skin cancer lesion treatment. In order to establish an adequate level of quality of treatment, a risk analysis of the Esteya treatment process has been done, following the methodology proposed by the TG-100 guidelines of the American Association of Physicists in Medicine (AAPM). Material and methods: A multidisciplinary team familiar with the treatment process was formed. This team developed a process map (PM) outlining the stages, through which a patient passed when subjected to the Esteya treatment. They identified potential failure modes (FM) and each individual FM…

research product

High-dose-rate brachytherapy boost for prostate cancer: Analysis of dose-volume histogram parameters for predicting late, rectal toxicity

PURPOSE: To determine the significance of dose-volume histogram parameters for predicting late rectal toxicity (LRT) after single-fraction high-dose-rate brachytherapy (HDRBT) boost and external beam radiotherapy (EBRT) in prostate cancer. MATERIALS AND METHODS: Three hundred patients with intermediate- or high-risk prostate cancer were included between August 2010 and March 2015. Treatment comprised a single-fraction HDRBT boost of 15.0 Gy plus EBRT (46.0 Gy delivered in 23 fractions) or an HDRBT boost of 9.5 Gy plus EBRT (60.0 Gy delivered in 30 fractions) if the seminal vesicles were infiltrated using real-time transrectal ultrasound-based planning. LRT was evaluated every 3 months after…

research product

Fetal dose measurements and shielding efficiency assessment in a custom setup of (192)Ir brachytherapy for a pregnant woman with breast cancer.

To assess the radiation dose to the fetus of a pregnant patient undergoing high-dose-rate (HDR) (192)Ir interstitial breast brachytherapy, and to design a new patient setup and lead shielding technique that minimizes the fetal dose.Radiochromic films were placed between the slices of an anthropomorphic phantom modeling the patient. The pregnant woman was seated in a chair with the breast over a table and inside a leaded box. Dose variation as a function of distance from the implant volume as well as dose homogeneity within a representative slice of the fetal position was evaluated without and with shielding.With shielding, the peripheral dose after a complete treatment ranged from 50 cGy at…

research product