Simultaneous detection of three antiviral and four antibiotic compounds in source-separated urine with liquid chromatography
An analytical method for the simultaneous screening of three antiviral agents (nevirapine, zidovudine, lamivudine), four antibiotics (sulfamethoxazole, trimethoprim, ciprofloxacin, rifampicin) and one reference compound (carbamazepine) in human urine was developed. Separation was achieved with a Kinetex XB-C18 (75 × 4.6 mm, 2.6 μm) column after the extraction of pharmaceuticals from urine with SPE. Gradient elution with a mobile phase consisting of acetonitrile and 10 mM KH2 PO4 (pH 2.5), and diode array detection with monitoring at 210 and 264 nm was applied. The developed method was validated in terms of selectivity, linearity, stability and sensitivity. Repeatability (n = 3) and between-…