Assessment of haemoglobin measurement by several methods - blood gas analyser, capillary and venous HemoCue® , non-invasive spectrophotometry and laboratory assay - in term and preterm infants
A laboratory haematology analyser is the gold standard for measuring haemoglobin concentration but has disadvantages, especially in neonates. This study compared alternative blood-sparing and non-invasive methods of haemoglobin concentration measurement with the gold standard. Haemoglobin concentrations were measured using a laboratory haematology analyser (reference method), blood gas analyser, HemoCue® using venous and capillary blood samples and a newly developed non-invasive sensor for neonates < 3 kg. A total of 63 measurements were performed. Body weight (2190 (1820-2520 [967-4450]) g) and haemoglobin concentration (12.3 (10.6-15.2 [8.2-20.5]) g.dl-1 ) varied widely. Bias/limits of ag…