0000000000287612

AUTHOR

Manuel Marcos-robles

showing 1 related works from this author

Optical Characterization Method for Tilted or Decentered Intraocular Lenses

2016

Purpose To test and validate a method for measuring the optical quality and optical power of monofocal intraocular lenses (IOLs) in the presence of a tilt or a decentration as well as its optical power. Methods The experimental system consists of an artificial eye (wet-cell) and a commercial aberrometer with a Shack-Hartmann sensor. Optical image quality parameters such as point spread function (PSF), modulation transfer function (MTF), and the simulations of the retinal image of an extended object are computed from the in vitro wavefront data of an IOL. Repeatability and reproducibility of the system are tested. Mathematical and ray-tracing simulations are used to ascertain the precision a…

Point spread functionOptics and PhotonicsAccuracy and precisionMaterials sciencePseudophakiaOptical powerArtificial Lens Implant Migration01 natural sciences010309 optics03 medical and health sciences0302 clinical medicineOpticsLens Implantation IntraocularOptical transfer function0103 physical sciencesHumansLenses IntraocularWavefrontReproducibilitybusiness.industryAberrometryReproducibility of ResultsRepeatabilityModels TheoreticalOphthalmologyTilt (optics)030221 ophthalmology & optometrybusinessOptometryOptometry and Vision Science
researchProduct