ICD Shock Reduction by Subthreshold Pacing Introduction Implantable cardioverter defibrillators (ICD) may have the capacity to provoke or worsen ventricular tachyarrhythmias (VT). It has been reported that ICD shocks by itself can increase mortality. This study aimed to determine the role of back-up pacing-induced VT (PIT) in the overall ICD shock burden by avoiding pause-related ventricular back-up pacing. Methods and Results A population of 550 single-chamber ICD patients was studied. Of them, 17 (3%, 69 ± 16 years, 14 male) patients had documented episodes of PIT. A total of 431 VT episodes were documented including 89 (21%) due to PIT. In 3 patients, VT events were exclusively PITs. Aft…