0000000000297537
AUTHOR
A Otal
SU-GG-T-70: A Dosimetric Study of the MicroSelectron HDR Ir-192 Brachytherapy Source
Purpose: Recently the manufacturer of the mHDR‐v2 HDR Ir‐192 brachytherapy source clarified design changes that may alter the existing dosimetric data for this source. The object of this study was to obtain new dose rate tables following the TG‐43 formalism and to provide these data in close proximity to the source with high spatial resolution including the electron dosimetric contributions from Ir‐192 disintegration. Methods and Material: To study the source we used three different Monte Carlo codes: MCNP5, GEANT4, and Penelope. The source was located inside a 40 cm radius water phantom. Dose and kerma were obtained using 0.1 mm thick voxels to provide high‐resolution dosimetry near the so…
A practical MRI-based reconstruction method for a new endocavitary and interstitial gynaecological template
Purpose: There are perineal templates for interstitial implants such as MUPIT and Syed applicators. Their limitations are the intracavitary component deficit and the necessity to use computed tomography (CT) for treatment planning since both applicators are non-magnetic resonance imaging (MRI) compatibles. To overcome these problems, a new template named Template Benidorm (TB) has been recently developed. Titanium needles are usually reconstructed based on their own artifacts, mainly in T1-weighted sequence, using the void on the tip as the needle tip position. Nevertheless, patient tissues surrounding the needles present heterogeneities that complicate the accurate identification of these …