A. Friesenecker

Realization of administration unit for3He with gas recycling

Hyperpolarized (HP) noble gases (3He,129Xe) are used for MR-imaging of the lung. In the majority of case the HP gas is filled in Tedlarbags and directly inhaled by the patients. Starting from an earlier pilot device, an administration unit was built respectively to the Medical Devices Law to administer patients HP noble gas boli in defined quantities and at a predefined time during inspiration with high reproducibility and reliability without reducing MR-quality. The patient's airflows are monitored and recorded. It is possible to use gas admixtures, measure the polarization on-line and collect the exhaled gas for later recycling. The first images with healthy volunteers were taken with thi…

research product

Application unit for the administration of contrast gases for pulmonary magnetic resonance imaging: optimization of ventilation distribution for3He-MRI

Purpose MRI of lung airspaces using gases with MR-active nuclei (3He, 129Xe, and 19F) is an important area of research in pulmonary imaging. The volume-controlled administration of gas mixtures is important for obtaining quantitative information from MR images. State-of-the-art gas administration using plastic bags (PBs) does not allow for a precise determination of both the volume and timing of a 3He bolus. Methods A novel application unit (AU) was built according to the requirements of the German medical devices law. Integrated spirometers enable the monitoring of the inhaled gas flow. The device is particularly suited for hyperpolarized (HP) gases (e.g., storage and administration with m…