0000000000370981
AUTHOR
Ana Ros García
Impact of crystal quality, geometry and surface finish for 3D impact position measurements in gamma ray detection systems
We have studied several ways of optimizing the 3D impact position measurements for gamma ray imaging detectors based on continuous monolithic large sized and thick scintillation crystals. The readout of such detectors is performed with position sensitive photo multipliers and with a depth of interaction (DOI) enhanced charge division readout. We have studied the effect of the crystal truncation angle, the reflective covering and the absorption and scattering mean free paths of the crystal. In particular, we have simulated and measured the impact on the energy resolution and the 3D spatial resolution. The results show that by optimization of the accessible crystal parameters we can enhance t…
Optimización de cristales centelleadores para la determinación de la DOI en tomografía de rayos gamma
161 páginas. Tesis Doctoral del Departamento de Física Atómica, Molecular y Nuclear de la Universidad de Valencia y del Instituto de Física Corpuscular (IFIC).