Internal ion impact ionization for Fourier-transform ion cyclotron resonance
A general scheme for ionization and fragmentation of ions in Fourier-transform ion cyclotron resonance mass spectrometry (FT-ICRMS) is introduced. The technique is based upon excitation of high-frequency (i.e. lowmass) primary ions (e.g. N, Al+, H2O+·) which may be generated by standard ionization methods (e.g. electron ionization or laser desorption) inside the trap. Generation of the primary ions is followed by excitation of their cyclotron motion to high translational energy to serve as projectile ions. The projectiles serve two functions: to ionize neutral gas atoms and/or molecules (secondary ions) within the trap internal ion impact ionization (IIII), and to collisionally activate the…