Influence of Implant Connection, Abutment Design and Screw Insertion Torque on Implant-Abutment Misfit
Background: An accurate fit at the implant-abutment interface is an important factor to avoid biological and mechanical complications. The aim of this study was to evaluate the marginal misfit at the implant-abutment interface on external and Morse taper connection, with straight and angulated abutments under different insertion torque loads. Materials and Methods: A total of 120 implants were used, 60 with external connection (EC) and 60 with Morse taper connection (IC). Straight (SA) (n = 60) and angulated abutments (AA) (n = 60) were randomly screwed to each connection at different torque levels (n = 10 each): 10, 20 and 30 Ncm. All specimens were subjected to thermal and cyclic loading …