0000000000419386

AUTHOR

Esko Kyrö

Operational considerations to improve total ozone measurements with a Microtops II ozone monitor

A Microtops II "ozone monitor" with UV channels centered at 305.5, 312.5, and 320 nm has been used routinely in six experimental campaigns carried out in several geographic locations and seasons, covering latitudes from 35 to 68° N during the last ten years (2001–2011). The total ozone content is retrieved by Microtops II by using different combinations (Channel I, 305.5/312.5 nm; Channel II, 312.5/320 nm; and Channel III, 305.5/312.5/320 nm) of the signals at the three ultraviolet wavelengths. The long-term performance of the total ozone content determination has been studied taking into account the sensitivities to the calibration, airmass, temperature and aerosols. When a calibration was…

research product

In search of traceability: two decades of calibrated Brewer UV measurements in Sodankylä and Jokioinen

Abstract. The two Brewer spectrophotometers of the Finnish Meteorological Institute at Jokioinen and Sodankylä have been operated according to the highest levels of the WMO∕GAW (World Meteorological Organization∕Global Atmosphere Watch) recommendations with rigorous quality control and quality assurance. The calibration of the instruments is based on annual recalibrations of primary standard lamps in the VTT MIKES Metrology National Standards Laboratory in Finland and an exhaustive measurement program with measurements of standard and working lamps in the on-site optical laboratories. Over the years, the maintenance of the calibration has produced data sets of approximately 2000 lamp scans …

research product