0000000000429289
AUTHOR
T. García Martinez
EP-1168: RapidArc independent monitor unit calculations with Diamond. Practical considerations
SU-E-T-218: Octavius 4D: Commissioning and Clinical Implementation of a New Device for VMAT Verifications
Purpose: To commission a new device for pre‐treatment patient specific verification of VMAT plans, which is capable of reconstructing 3D dose with high resolution. Methods: Octavius 4D system (PTW) consists of an ion chamber array embedded in a cylindrical phantom which, assisted by an inclinometer, rotates synchronously with the gantry. For VMAT plans, it measures planar dose distributions as a function of gantry angle in order to compute the resulting 3D dose distribution. There are two options for the ion chamber array: Octavius 729 (general purpose) and Octavius 1000 SRS (small field size, high resolution). We have used three linacs: two Varian Clinac iX and one TrueBeam with MLC HD 120…