0000000000442878
AUTHOR
L. Geraci
Analysis of signal-to-noise ratio for a 2-channels coil developed to enable transcranial Magnetic Resonance-guided Focused Ultrasound Surgery (tcMRgFUS) with 1.5 T MRI scanners
Transcranial Magnetic Resonance-guided Focused Ultrasound Surgery (tcMRgFUS) at 1.5T: this is how we do it
Aims and objectives Methods and materials Results Conclusion Personal information References
Trans-cranial MRI-guided Focused Ultrasound Surgery (tcMRgFUS): preliminary Italian (and world-first) experience at 1,5 Tesla.
PURPOSE: We present our preliminary results achieved with the first Italian installation of a trans-cranial MRI-guided Focused Ultrasound Surgery (tcMRgFUS) certified system for functional neurosurgery. Moreover, to our knowledge, this is the world-first tcMRgFUS system ever installed on a 1.5T MRI unit. Technical issues faced to achieve a safe and effective treatment will be discussed focusing on MR high-resolution live imaging and thermometry sequences optimization. METHOD AND MATERIALS: Patient enrollment was based on indication for functional neurosurgery and evidence of medication-refractory disease; a detailed medical history has been collected together with a complete clinical examin…