0000000000542219

AUTHOR

Néstor Ferrando

Position sensitive scintillator based detector improvements by means of an integrated front-end

PESIC is an integrated front-end for multianode photomultiplier based nuclear imaging devices. Its architecture has been designed to improve position sensitive detectors behavior by equalizing its response over its whole area. Its preamplying stage introduces two main benefits: digitally programmable gain adjustment for every photomultiplier output, and isolation from other front-end electronics by means of current buffers. This last feature allows to use different types of photomultipliers and optimizes front-end deadtime, reducing impact position dependent output delay. PESIC also includes an indirect measurement of the depth of interaction of the gamma ray inside the scintillator crystal…

research product

Maximum likelihood positioning for gamma-ray imaging detectors with depth of interaction measurement

Abstract The center of gravity algorithm leads to strong artifacts for gamma-ray imaging detectors that are based on monolithic scintillation crystals and position sensitive photo-detectors. This is a consequence of using the centroids as position estimates. The fact that charge division circuits can also be used to compute the standard deviation of the scintillation light distribution opens a way out of this drawback. We studied the feasibility of maximum likelihood estimation for computing the true gamma-ray photo-conversion position from the centroids and the standard deviation of the light distribution. The method was evaluated on a test detector that consists of the position sensitive …

research product