Dose calculation in biological samples in a mixed neutron-gamma field at the TRIGA reactor of the University of Mainz
To establish Boron Neutron Capture Therapy (BNCT) for non-resectable liver metastases and for in vitro experiments at the TRIGA Mark II reactor at the University of Mainz, Germany, it is necessary to have a reliable dose monitoring system. The in vitro experiments are used to determine the relative biological effectiveness (RBE) of liver and cancer cells in our mixed neutron and gamma fi eld. We work with alanine detectors in combination with Monte Carlo simulations, where we can measure and characterize the dose. To verify our calculations we perform neutron fl ux measurements using gold foil activation and pin-diodes . Material and methods . When L- α -alanine is irradiated with ionizing …
Dose determination using alanine detectors in a mixed neutron and gamma field for boron neutron capture therapy of liver malignancies
IntroductionBoron Neutron Capture Therapy for liver malignancies is being investigated at the University of Mainz. One important aim is the set-up of a reliable dosimetry system. Alanine dosimeters have previously been applied for dosimetry of mixed radiation fields in antiproton therapy, and may be suitable for measurements in mixed neutron and gamma fields.Materials and MethodsTwo experiments have been carried out in the thermal column of the TRIGA Mark II reactor at the University of Mainz. Alanine dosimeters have been irradiated in a phantom and in liver tissue.ResultsFor the interpretation and prediction of the dose for each pellet, beside the results of the measurements, calculations …
The alanine detector in BNCT dosimetry: Dose response in thermal and epithermal neutron fields
Purpose: The response of alanine solid state dosimeters to ionizing radiation strongly depends on particle type and energy. Due to nuclear interactions, neutron fields usually also consist of secondary particles such as photons and protons of diverse energies. Various experiments have been carried out in three different neutron beams to explore the alanine dose response behavior and to validate model predictions. Additionally, application in medical neutron fields for boron neutron capture therapy is discussed. Methods: Alanine detectors have been irradiated in the thermal neutron field of the research reactor TRIGA Mainz, Germany, in five experimental conditions, generating different secon…
Evaluation of the relative thermoluminescence efficiency of LiF:Mg,Ti and LiF:Mg,Cu,P TL detectors to low-energy heavy ions
Abstract The most popular types of LiF-based thermoluminescent (TL) detectors, LiF:Mg,Ti (MTS-N) and LiF:Mg,Cu,P (MCP-N), have been investigated, with respect to their dose (fluence) response and relative TL efficiency to different ion species. The detectors were irradiated using the nitrogen, iron, krypton and xenon ion beams, at energies ranging from 5.0 to 9.3 MeV/n. Supra- and sublinear response was found, for the MTS-N and MCP-N, respectively, similarly as observed for γ-rays. However, the level of nonlinearity of response of studied detectors is strongly reduced by increasing values of the ion ionization density (no supralinearity for Xe ions, for MTS-N, within calculated uncertaintie…