Targeting Moderate and Severe Male Stress Urinary Incontinence With Adjustable Male Slings and the Perineal Artificial Urinary Sphincter: Focus on Perioperative Complications and Device Explantations.
Purpose: To analyze perioperative complications and postoperative explantation rates for selected readjustable male sling systems and the perineal single-cuff artificial urinary sphincter (AUS) in a large, contemporary, multi-institutional patient cohort. Methods: Two hundred eighty-two male patients who underwent implantation between 2010 and 2012 in 13 participating institutions were included in the study (n = 127 adjustable male sling [n = 95 Argus classic, n = 32 Argus T], n = 155 AUS). Perioperative characteristics and postoperative complications were analyzed. The explantation rates of the respective devices were assessed using the Fisher exact test and the Mann-Whitney U-test. A Kapl…