Oxygen microoptodes: a new tool for oxygen measurements in aquatic animal ecology
We describe two applications of a recently introduced system for very precise, continuous measurement of water oxygen saturation. Oxygen microoptodes (based on the dynamic fluorescence quenching principle) with a tip diameter of similar to50 mum, an eight-channel optode array, an intermittent flow system, and online data registration were used to perform two types of experiments. The metabolic activity of Antarctic invertebrates (sponges and scallops) was estimated in respiration experiments, and, secondly, oxygen saturation inside living sponge tissue was determined in different flow regimes. Even in long-term experiments (several days) no drift was detectable in between calibrations. Data…