Ventilator Settings to Avoid Nuisance Alarms During Mouthpiece Ventilation
BACKGROUND: A recent study found that activation of disconnection and low-pressure alarms is common during mouthpiece ventilation and may represent a major limitation to its use. The aim of this bench study was: (1) to investigate the technical aspects that can influence the setting of the ventilator during mouthpiece ventilation and (2) to provide a practical setting strategy to avoid the alarm activation. METHODS: Eight life-support ventilators able to deliver volume controlled ventilation were tested in a bench study using a single-limb non-vented circuit configuration connected to a standard mouthpiece. Disconnection and apnea alarm were turned off or set at the least sensitive setting.…