Evaluation of cut quality using the Amadeus microkeratome with different settings.
Abstract Purpose: To evaluate the quality of keratectomy specimens created with the Amadeus® microkeratome (AMO) using scanning electron microscopy (SEM). Setting: Johannes Gutenberg-University Hospital, Department of Ophthalmology, Mainz, and the Department of Anatomy, Justus Liebig-University, Giessen, Germany. Methods: Corneal cuts were performed in freshly enucleated pig eyes using the Amadeus microkeratome with 9 combinations of oscillation rate and head-advance speed. For the cutting trials, oscillation rates of 8000, 13000, and 18000 rpm and head-advance speeds of 1.0, 2.5, and 4.0 mm/s were chosen. Fifty-four eyes were included, resulting in 9 groups of 6 eyes for each configuration…