6533b829fe1ef96bd128af43
RESEARCH PRODUCT
Evaluation of cut quality using the Amadeus microkeratome with different settings.
H. Burkhard DickStefanie SchäferMana Tehranisubject
Quality Controlbusiness.industryOscillationSwinemedicine.medical_treatmentCorneal StromaKeratomileusis Laser In SituKeratomileusisSurface smoothnessSensory SystemsSurgical FlapsOphthalmologyQuality (physics)OpticsEvaluation Studies as TopicMicrokeratomeRefractive surgerySurface roughnessmedicineMicroscopy Electron ScanningAnimalsSurgerybusinessMathematicsdescription
Abstract Purpose: To evaluate the quality of keratectomy specimens created with the Amadeus® microkeratome (AMO) using scanning electron microscopy (SEM). Setting: Johannes Gutenberg-University Hospital, Department of Ophthalmology, Mainz, and the Department of Anatomy, Justus Liebig-University, Giessen, Germany. Methods: Corneal cuts were performed in freshly enucleated pig eyes using the Amadeus microkeratome with 9 combinations of oscillation rate and head-advance speed. For the cutting trials, oscillation rates of 8000, 13000, and 18000 rpm and head-advance speeds of 1.0, 2.5, and 4.0 mm/s were chosen. Fifty-four eyes were included, resulting in 9 groups of 6 eyes for each configuration. The surfaces and edges of the flaps were examined using SEM. Results: At fixed oscillation rates, an increase in head-advance speed led to lower quality cuts, higher surface roughness, and irregular cut edges. At fixed head-advance speeds, an increase in oscillation rates improved the cut quality, resulting in smoother surface characteristics. Conclusions: Smooth and regular surfaces and cut edges can be achieved by choosing high oscillation rates and low head-advance speeds. Microkeratomes that enable the user to adjust the oscillation rate and head-advance speed can be used to maximize surface smoothness and improve customized refractive surgery.
year | journal | country | edition | language |
---|---|---|---|---|
2004-03-04 | Journal of cataract and refractive surgery |