Sensitive determination of polycyclic aromatic hydrocarbons in water samples using monolithic capillary solid-phase extraction and on-line thermal desorption prior to gas chromatography-mass spectrometry.
A methacrylate-based monolithic capillary column has been evaluated for the preconcentration of polycyclic aromatic hydrocarbons (PAHs) from environmental water samples. For this purpose, the monolyte was in situ synthesized in a 6cm×0.32mm id fused-silica capillary. The microextraction unit was fitted to a micro-HPLC pump to pass 10mL of sample. The isolated pollutants were eluted by means of 10μL of methanol, the organic phase being directly collected in a specific interface that can be fitted to the injection port of the gas chromatograph without modification. The interface allows the on-line thermal desorption of the PAHs, avoiding the dilution and providing enough sensitivity to reach …