Multipumping Nitrite Determination in Exhaled Breath Condensate

2006

Abstract A flow system based on multicommutation is proposed for the rapid, clean, and inexpensive determination of nitrites in small volumes of breath condensates. The procedure exploits the colorimetric detection of nitrite with the Griess reagent [0.03% naphthylethylene diamine dihydrochloride (NED), 0.5% sulpfhanilamide, and 3.0% phosphoric acid] in acidic medium at 540 nm correcting the variations of the baseline with measurements at 424 nm. The flow system was designed with a set of solenoid micropumps to minimize sample and reagent consumption and waste generation. The detection limit was estimated as 3.8 ng mL−1 (99.7% confidence level) with a linear response ranging up to 500 ng mL…