SU-E-T-176: Commissiong and Initial Clinical Experience with Dosimetry Check, a Commercial Software for in Vivo Volumetric Dosimetry

Purpose: The aim of this work is to study the differences of reference point doses between the commercial EPID based dosimetry software Dosimetry Check (DC) and TPS to establish an accuracy level of the system, to evaluate its use in clinical routine and to obtain results for the first patients. Methods: We used DC v.3.8 (Math Resolutions), two Varian Clinac iX accelerators equipped with EPID aS1000 and Eclipse v.10.0 with AAA algorithm. Several plans with and without air gap were generated over the phantoms: MP1 Water Tank and Solid Octavius 4D (PTW). 5 head and neck, 3 lung and 4 prostate cases were selected for clinical evaluation. We compared the calculated dose at the isocenter and in …

research product

SU-E-T-218: Octavius 4D: Commissioning and Clinical Implementation of a New Device for VMAT Verifications

Purpose: To commission a new device for pre‐treatment patient specific verification of VMAT plans, which is capable of reconstructing 3D dose with high resolution. Methods: Octavius 4D system (PTW) consists of an ion chamber array embedded in a cylindrical phantom which, assisted by an inclinometer, rotates synchronously with the gantry. For VMAT plans, it measures planar dose distributions as a function of gantry angle in order to compute the resulting 3D dose distribution. There are two options for the ion chamber array: Octavius 729 (general purpose) and Octavius 1000 SRS (small field size, high resolution). We have used three linacs: two Varian Clinac iX and one TrueBeam with MLC HD 120…