Effective broadband refractive index retrieval by a white light optical particle counter.
A new approach to retrieve the effective broadband refractive indices (nbroad,eff) of aerosol particles by a white light aerosol spectrometer (WELAS) optical particle counter (OPC) is presented. Using a tandem differential mobility analyzer (DMA)-OPC system, the nbroad,eff are obtained for both laboratory and field applications. This method was tested in the laboratory using substances with a wide range of optical properties. With the obtained nbroad,eff, WELAS aerosol size distributions can be corrected. Therefore, this method can be used for instrument calibration in both laboratory and field measurements. The retrieved effective broadband refractive indices for the scattering aerosols we…