An in-tube SPME device for the selective determination of chlorophyll a in aquatic systems.
We report a new device for the estimation of the content of chlorophyll a pigment in water samples as an indicator of water quality. The extraction of the pigment from water was also optimized. 10 mL of water was filtered through a nylon filter (45m pore size and 13 mm of diameter), after the chlorophylls were dissolved by immersing the filter in 1 mL of a low non-hazardous solvent as ethanol. An in-valve in-tube SPME device coupled to capillary liquid chromatography with diode array detection was designed. A capillary column of 70 cm in length (0.32 mm i.d. coated with 5% diphenyl–95% polydimethylsiloxane, 3m coating thickness) was used as the loop of the injection valve for preconcentrati…