0000000000794211

AUTHOR

B. Juste

Improving pediatric radiation dose management using Agfa digital radiography DICOM header information

Dose reduction in pediatric explorations is especially important because of children radiation sensitivity. According to this, with the aim of saving radiation exposure in future clinical practice, we have developed a technique to control delivered dose in pediatric radiographic exams. To that, a computer science program has been developed to calculate entrance skin dose (ESD) provided by AGFA radiology digital system, using the "lgM" parameter exported from Dicom files. ESD values are compared with dose limits established in regulations to detect if children are being exposed to excessive amounts of radiation during their explorations.

research product

Comparison between transmission and scattering spectrum reconstruction methods based on EPID images.

Numerous improved physics-based methods for Linac photon spectra reconstruction have been published; some of them are based on transmission data analysis and others on scattering data. In this work, the two spectrum unfolding approaches are compared in order to experimentally validate its robustness and to determine which is the optimal methodology for application on a clinical quality assurance routine. Both studied methods are based on EPID images generated when the incident photon beam impinges onto plastic blocks. The distribution of transmitted/scatter radiation produced by this object centered at the beam field size was measured. Measurements were performed using a 6 MeV photon beam p…

research product