0000000000799388

AUTHOR

Francesc Borrull

0000-0003-2718-9336

showing 2 related works from this author

Validation of a confirmatory method for the determination of macrolides in liver and kidney animal tissues in accordance with the European Union regu…

2006

Abstract This study proposes a simple multiresidue liquid chromatography–diode array detector (LC–DAD) method capable of determining seven macrolide antibiotics in samples of liver and kidney animals at concentrations lower than those allowed by current legislation. Samples were prepared by homogenizing the tissue with EDTA–McIlvaine's buffer and extracted with an Oasis HLB cartridge. The consumption of organic solvent during extraction was minimum. The analytes were detected by LC–DAD and also by liquid chromatography–mass spectrometry with electrospray ionization (LC–(ESI)MS). The method was specific, stable and robust enough for the required purposes. The DAD method was validated in acco…

AnalyteChromatographyChemistryElectrospray ionizationOrganic ChemistryReproducibility of ResultsGeneral MedicineKidneySensitivity and SpecificityBiochemistryHigh-performance liquid chromatographyAnti-Bacterial AgentsAnalytical ChemistryLiverCalibrationAnimalsmedia_common.cataloged_instanceSample preparationEuropean UnionMacrolidesSolid phase extractionEuropean unionQuantitative analysis (chemistry)Antibacterial agentmedia_commonJournal of Chromatography A
researchProduct

Determination of macrolide antibiotics in meat and fish using pressurized liquid extraction and liquid chromatography–mass spectrometry

2008

We developed a method for determining the quantities of seven macrolide antibiotics in meat and fish by using pressurized liquid extraction (PLE) and liquid chromatography-mass spectrometry with electrospray ionization (LC-(ESI)MS). The PLE was optimized with regard to solvents, temperature, pressure, extraction time and number of cycles. The optimum conditions were: methanol as the extraction solvent; a temperature of 80 degrees C; a pressure of 1500psi; an extraction time of 15min; 2 cycles; a flush volume of 150% and a purge time of 300s. All recoveries for macrolide antibiotics were over 77% at 200mug/kg, except for erythromycin, which was 58%. The repeatability and reproducibility on d…

Spectrometry Mass Electrospray IonizationElectrosprayMeatChromatographySwineChemistryElectrospray ionizationOrganic ChemistryExtraction (chemistry)FishesAnalytic Sample Preparation MethodsAnalytic Sample Preparation MethodsGeneral MedicineMass spectrometryBiochemistryHigh-performance liquid chromatographyDrug ResiduesAnti-Bacterial AgentsAnalytical ChemistryLiquid chromatography–mass spectrometryAnimalsCattleChickensChromatography LiquidAntibacterial agentJournal of Chromatography A
researchProduct