In vitro retention capacity of two overdenture attachment systems: Locator® and Equator®
Background It is necessary to know the in vitro behavior of different attachment systems to be used clinically. The evolution of retention capacity over 10 years (14,600 insertion/de-insertion cycles) was determined in vitro, evaluating two overdenture attachment systems (Locator® and OT Equator®). Material and methods The study used an implant replica compatible with the abutments of both systems. 10 Locator® and 10 OT Equator® attachments were screwed to the abutments. Nylon inserts were attached and tested, subjecting them to 14,600 insertion and de-insertion cycles (representing 10 years functional life) in axial direction. The universal test machine crosshead speed was 50 mm/min with a…