The AVL-mode: a safe closed loop algorithm for ventilation during total intravenous anesthesia.
The Adaptive Lung Ventilation Controller (ALV-Controller) represents a new approach to closed loop control of ventilation. It is based on a pressure controlled ventilation mode. Adaptive lung ventilation signifies automatic breath by breath adaptation of breathing patterns to the lung mechanics of an individual patient. The specific goals are to minimize work of breathing, to maintain a preset alveolar ventilation and to prevent the occurrence of intrinsic PEEP. We ventilated 5 patients undergoing major abdominal procedures using ALV. ALV was tolerated well in all patients. Alveolar ventilation was preset between 5500 and 6500 ml/min. Serial dead space (Vds) and respiratory time constant (r…