Computational Fluid Dynamics (CFD) and Finite Element Analysis (FEM) of a Customized Stent-Graft for Endovascular (EVAR) Treatment of Abdominal Aortic Aneurism (AAA)
Background: The treatment of abdominal aortic aneurysm (AAA) is today commonly treated by inserting a stent-graft by the endovascular route, without resorting to open surgery. However, some clinical cases do not allow this less invasive approach, meaning that the stent-graft cannot be inserted and open surgery is used. Methods: In the study, we propose a fluid–structure interaction (FSI) analysis of an aneurysmatic aorta that could not be treated with Endovascular Aneurysm Repair (EVAR). The vessel is reconstructed through segmentation from CT scans and subsequently modeled on CAD software to create the surface and thickness of the vessel itself. Subsequently, we proceeded to carry out Comp…