GEC-ESTRO ACROP recommendations on calibration and traceability of LE-LDR photon-emitting brachytherapy sources at the hospital level
Prostate brachytherapy treatment using permanent implantation of low-energy (LE) low-dose rate (LDR) sources is successfully and widely applied in Europe. In addition, seeds are used in other tumour sites, such as ophthalmic tumours, implanted temporarily. The calibration issues for LE-LDR photon emitting sources are specific and different from other sources used in brachytherapy. In this report, the BRAPHYQS (BRAchytherapy PHYsics Quality assurance System) working group of GEC-ESTRO, has developed the present recommendations to assure harmonized and high-quality seed calibration in European clinics. There are practical aspects for which a clarification/procedure is needed, including aspect…