Long-Term Outcome with New Generation Prostheses in Patients Undergoing Transcatheter Aortic Valve Replacement
The aim of this study was to compare patients with transcatheter aortic valve replacement (TAVR) receiving new generation prostheses SAPIEN 3 (S3, Edwards Lifesc.) and Evolut R (ER, Medtronic Inc.) in terms of periprocedural and long-term outcome. Our retrospective, single-center analysis included 359 consecutive patients with severe aortic stenosis who underwent TAVR with S3 or ER from 2014–2016 (mean age 82 ± 7 years, 47% male, mean EuroSCORE II 8.0 ± 8%, mean follow-up 3.8 years). Device Success was equal (S3 93.0% vs. ER 92.4%, p = 0.812). We report a 30-day mortality of 2.8% in the S3 group, and 2.1% in the ER group (p = 0.674). There was no difference in stroke, conversion to open sur…