Early results of a low-profile stent-graft for thoracic endovascular aortic repair
Purpose To assess outcomes of a low-profile thoracic stent-graft in the treatment of thoracic aortic pathologies. Methods A retrospective analysis of all consecutive patients with aortic thoracic pathologies treated with the RelayPro device in two university hospitals between October 2018 and July 2019. Results 23 patients (65% men; mean age 63.4 ± 15 years) were treated. Pathologies included aortic dissections (n = 10), 5 residual type A (22%) and 5 type B (22%), 6 degenerative aortic aneurysms (26%), 4 penetrating aortic ulcers (17%), and aortic erosion, intramural hematoma and aortic rupture (n = 1 and 4% in each case). Two cases (9%) were emergent and two urgent. Proximal landing was a…