Dosimetric Characteristics of a New Unit for Electronic Skin Brachytherapy
SU-E-T-343: Valencia Applicator Commissioning Using a Micro-Chamber Array
Purpose: In the commissioning and QA of surface isotope-based applicators, source-indexer distance (SID) has a great influence in the flatness, symmetry and output. To these purposes, methods described in the literature are the use of a special insert at the entrance of dwell chamber or radiochromic films. Here we present the experience with a micro-chamber array to perform the commissioning and QA of Valencia applicators. Methods: Valencia applicators have been used, the classic and the new extra-shielded version. A micro-chamber array has been employed, 1000 SRS (PTW), with 977 liquid filled, 2.3×2.3×0.5 mm3 sized ion chambers covering 11×11 cm2, which spacing is 2.5 mm in the central 5.5…