In vitro comparison of platinum-iridium and gold tip electrodes: lesion depth in 4 mm, 8 mm, and irrigated-tip radiofrequency ablation catheters
Aims We compared a newly developed irrigated gold tip electrode ablation catheter and a gold tip 4 and 8 mm catheter with the corresponding platinum–iridium (Pt) tip catheters in an in vitro setting. Methods and results In a flow chamber simulating physiological flow conditions, radiofrequency catheter ablation was performed on tissue samples of porcine endomyocardium and liver. Lesion depth, energy and temperature delivery, and popping frequency were determined. Two hundred and fifty-three ablations were conducted. Four and eight millimetre, gold tip electrode catheters produced significantly deeper lesions compared with the Pt tip electrode (liver 4 mm: 4.67 ± 1.7 vs. 2.9 ± 1.0 mm, P < 0.…