SU-E-T-509: DICOM Test Case Plans for Model-Based Dose Calculations Methods in Brachytherapy
Purpose: The TG‐186 report provides guidance to early adopters of model‐based dose calculation algorithms (MBDCAs) for brachytherapy. A charge of the AAPM‐ESTRO Working Group on MBDCA is to develop well‐defined test case plans, available as references for the software commissioning process to be performed by end‐users. The aim of this work is to develop test case plans for a generic HDR 192 Ir source alone and in combination with a vaginal cylinder applicator with 180° tungsten‐alloy shielding in a DICOM‐based water phantom. Methods: A DICOM CT dataset was created with a 30 cm diameter water sphere surrounded by air. The voxel size was 1.33×1.33×1.33 mm3 for evaluating absorbed dose rate (c…