Mixed-model of ANOVA for measurement reproducibility in proteomics
This work is a statistical analysis of reproducibility of a MALDI-TOF mass spectrometry experiment. Its aim is to evaluate measurement variability and compare peak intensities from two types of MALDI-TOF platforms. We compared and commented on the abilities of Principal Component Analysis and mixed-model analysis of variance to evaluate the biological variability and the technical variability of peak intensities in different patients. The properties and hypotheses of both methods are summarized and applied to spectra from plasma of patients with Hodgkin lymphoma. Principal Component Analysis checks rapidly the balance between the two variabilities; however, a mixed-model analysis of varianc…