Heated breathing tubes affect humidity output of circle absorber systems.
Abstract Study Objective: To verify whether the airway climate in circle systems can be improved with heated breathing tubes. Design: Randomized, controlled, prospective clinical study. Setting: Operating theater of the Department of Maxillofacial Surgery. Patients: 26 adult patients undergoing prolonged anesthesia. Interventions: A total of 26 prolonged anesthetics were conducted in adult patients using a minimal fresh gas flow rate (0.6 L/min) and silicon breathing tubes (16 mm internal diameter) containing a heated coil. Group 1 ( n = 10 patients) was the control group; breathing tubes were unheated. In Group 2 ( n = 10 patients), breathing tubes were heated to 30°C. In Group 3 ( n = 6 p…