Continuous Wave fNIRS with Silicon Photomultiplier
This work is focused on the development of a Continuous Wave (CW) NIRS integrated system with multi-wavelength LED sources between 700 and 950 nm and a Silicon Photomultiplier detector (SiPM) developed by STMicroelectronics. The Signal Noise Rate (SNR), measured placing the LEDs and an SiPM in a direct contact with the surface of a plastic phantom mimicking a real human head, results higher than the calculated minimum, required to detect small variation in the HbO2 and HHb concentration, till a source detector separation (SDS) of 6 cm.