Validation of Methylmercury Determinations in Aquatic Systems by Alkyl Derivatization Methods for GC Analysis Using ICP-IDMS
Isotope dilution mass spectrometry (IDMS), using an inductively coupled plasma quadrupole mass spectrometer (ICPMS) and a species-specific methylmercury spike was applied to validate the commonly used GC method for methylmercury (MeHg+) determination, which is based on the formation of volatile methylethylmercury by derivatization with NaBEt4. The spike compound, Me201Hg+, was synthesized by reaction of 201Hg-enriched mercury chloride with methylcobalamin. By analyzing different environmental aquatic samples, it was found that in most cases, transformation of MeHg+ into elemental mercury (Hg0) took place. From investigations of synthetic solutions, it could be followed that halide ions are …