0000000001247899
AUTHOR
Feyter Pj
Physiopathology of the aging heart
Influence of increasing convolution kernel filtering on plaque imaging with multislice CT using an ex-vivo model of coronary angiography
PURPOSE: To assess the variability in attenuation of coronary plaques with multislice CT-angiography (MSCT-CA) in an ex-vivo model with varying convolution kernels. MATERIALS AND METHODS: MSCT-CA (Sensation 16, Siemens) was performed in three ex-vivo left coronary arteries after instillation of contrast material solution (Iomeprol 400 mgI/ml, dilution: 1/80). The specimens were placed in oil to simulate epicardial fat. Scan parameters: slices 16/0.75 mm, rotation time 375 ms, feed/rotation 3.0 mm, mAs 500, slice thickness 1 mm, and FOV 50 mm. Datasets were reconstructed using 4 different kernels (B30f-smooth, B36f-medium smooth, B46f-medium, and B60f-sharp). Each scan was scored for the pre…