6533b7d1fe1ef96bd125c10a

RESEARCH PRODUCT

A purity monitoring system for liquid argon calorimeters

Peter BuchholzWolfgang WalkowiakD. WilkeA. HergesellF. HölldorferA. DahlhoffT. JungE. KuhnertC. SchwarzUli SchäferG. IelapiA. SchmidtChristian ZeitnitzY. SchuéM. AdamsM. HaarmannD. WeberKarl JakobsM. GramsD. FeidenV. GärtnerD. MederL. Köpke

subject

PhysicsNuclear and High Energy PhysicsResolution (mass spectrometry)Physics::Instrumentation and DetectorsImpurityDetectorElectron attachmentAnalytical chemistryLiquid argonHigh Energy Physics::ExperimentMonitoring systemPhysics::Chemical PhysicsInstrumentation

description

Abstract For liquid argon calorimeters electronegative impurities dissolved in the medium degrade the detector response and deteriorate the energy resolution, especially at high energies. A concept for a purity monitoring system for liquid argon calorimeters has been developed and is presented here. Special combined monitors of 241 Am- and 207 Bi-cells are used to monitor the concentration of impurities. The working principle as well as results from test measurements are discussed.

https://doi.org/10.1016/j.nima.2005.03.117