Extension of the dynamic range of flame atomic absorption spectrometry using flow injection analysis with variable-volume dilution chambers
H. El AzouziM. Y. Pérez-jordanAmparo SalvadorM. De La Guardiasubject
Flow injection analysisChemistryInstrumentationAnalytical chemistryStandard solutionAtomic and Molecular Physics and OpticsAnalytical ChemistryDilutionlaw.inventionStandard curveNebulizerVolume (thermodynamics)lawAtomic absorption spectroscopyInstrumentationSpectroscopydescription
Abstract A simple and inexpensive procedure is proposed for the extension of the dynamic range of flame atomic absorption spectrometry measurements using on-line dilution. The proposed methodology is based on the use of a manifold with two coupled dilution chambers and a zone injection system. The samples are prediluted in a closed system which includes a variable-volume mixing chamber (10–120 ml) and two injection valves. The samples are injected through one of these valves, and the other is employed to take 100 μl of prediluted samples which are then passed through a new dilution chamber (volume 1–10 ml) and aspirated by the nebulizer of the instrument. A third injection valve mounted in the last part of the manifold is used for the direct injection of diluted standard solutions. Various dilution factors are obtained, ranging from 2 to 130 000 times, thus extending the analytical range of copper determination to more than 100 000 mg l −1 .