Delayed-neutron spectroscopy with 3He spectrometers

H. OhmGünter HerrmannD.r. SlaughterF.m. NuhStanley G. PrussinW. RudolphKarl KratzH. Franz

subject

Range (particle radiation)Materials scienceSpectrometerIonization chamberMicrophonicsNeutronGeneral MedicineAtomic physicsNuclear ExperimentSpectroscopyDelayed neutronSpectral line

description

Abstract The characteristics of a 3 He ionization chamber for high-resolution spectroscopy of β-delayed neutrons have been examined. The detector response function is given in the energy range 0.019–2.77 MeV. A considerably improved energy resolution has been achieved by minimizing effects due to pulse summing, detection of scattered neutrons, and microphonics, and is most evident in the low-energy part of the spectra.

https://doi.org/10.1016/0029-554x(77)90116-1