6533b824fe1ef96bd127ff4d
RESEARCH PRODUCT
Theory of quadrupole detection fourier transform-ion cyclotron resonance
Lutz Schweikhardsubject
ChemistryCyclotronCyclotron resonanceAnalytical chemistryResonanceIon cyclotron resonance spectrometryFourier transform ion cyclotron resonancelaw.inventionsymbols.namesakeFourier transformlawQuadrupolesymbolsAtomic physicsSpectroscopyIon cyclotron resonancedescription
Abstract The theoretical basis of the recently introduced quadrupole detection Fourier transform-ion cyclotron resonance technique is developed by use of the concept of image charges induced in the detection electrodes. The appearance of resonances at frequencies 2ω+, 2ω−, (ω+ + ω−) and (ω+ − ω−) is explained. The new method is compared with the standard (dipole) detection scheme. Techniques are suggested allowing the simultaneous determination of all resonance frequencies ω+ (perturbed, or reduced, cyclotron frequency), ω− (magnetron frequency) and ωz (trapping frequency).
year | journal | country | edition | language |
---|---|---|---|---|
1991-07-01 | International Journal of Mass Spectrometry and Ion Processes |